Document Type : Original article
Authors
1 Department of Anesthesia & Critical Care, School of Medicine, Tehran University of Medical Sciences, Tehran, Iran
2 Tehran University of Medical Sciences, Tehran, Iran
3 Department of Thoracic Surgery, School of Medicine, Tehran University of Medical Sciences, Tehran, Iran
4 Department of Anesthesia & Critical Care, Tehran University of Medical Sciences
Abstract
Keywords
Abstract Background: Our goal was to compare two common methods including auscultation and Fiber-Optic Bronchoscopy (FOB) in confirming the correct placement of Double-Lumen Tube (DLT). Methods: Seventy six patients were enrolled. After DLT insertion, clinical verification was made by auscultation, then FOB was applied. At lateral decubitus, the position of DLT was rechecked by auscultation, and then by FOB. The incorrect position including malposition and misplacement were compared between two routine methods of auscultation and FOB. Results: After blind intubation, 21.1% of DLTs were considered to be in an incorrect position. Meanwhile, FOB showed that 46.1% of DLTs were not placed correctly. Among all the patients, 53.9% of DLTs were in the optimal position. Misplacement was diagnosed in 35.5% and malposition in 10.5% of the patients. After positioning to the lateral decubitus, using auscultation, we found that 5.3% of tubes were dislocated, but according to FOB, it was 10.5%. The agreement coefficient between auscultation and FOB was 0.42 in the supine position and 0.64 in lateral position. Conclusion: Our results showed that although FOB needs more time to check the position of DLT, it should be used to confirm the exact position of DLT. Keywords: Auscultation, Double-lumen tube, Fiberoptic bron- choscopy, Lateral decubitus, Thoracic surgery Introduction Lung isolation and one-lung ventilation using double-lumen tube are almost always required for thoracic surgery (1). The lung should be collapsed to let the surgeon proper visualization of the operation field and prevent contamination of the dependent lung and postoperative lower respiratory tract infection (2). Inappropriate Double-Lumen Tube (DLT) positioning can produce adverse events during One-Lung Ventilation (OLV). A malpositioned DLT will fail to allow the lung collapse, causing gas trapping during positive-pressure ventilation, or it may partially collapse the ventilated or dependent lung, producing hypoxemia. Therefore, proper positioning of Double-Lumen Tube (DLT) is critical (3,4). DLT movement, no matter how small, can cause difficulty in differential lung ventilation, produce hypoxemia and interfere with the surgical procedure (5,6). Change in body position from supine to lateral may lead to malposition of the DLT (7). Although majority of the anesthesiologists know that positioning can cause malposition of DLT, there are few studies evaluating the position of DLT before and after positioning. To avoid malposition of the DLT either during the insertion or after patient positioning, fiberoptic bronchoscopy has been recommended as a reliable method to evaluate DLT placement and lung separation (8). Despite these recommendations, some authors argue that bronchoscopy is costly, time-consuming, and not universally available, thus it should not be considered “routine” (9). Therefore, we designed this study to evaluate the malposition of DLT after positioning from supine to lateral using and comparing two methods of auscultation and fiberoptic bronchoscopy. Materials and Methods Following approval by the Tehran University of Medical Sciences School of Medicine ethics committee, 76 consecutive patients (ASA physical status II, III, and above 18 year-old) scheduled for thoracic surgery were enrolled in this cross-sectional study over an 8-month period. After explaining completely about the procedure, we obtained written informed consent. Patients with a difficult airway (Mallampati 3-4, or mouth opening<3 cm), airway anomaly, and anticipated use of right-sided DLT (like the patients with anomalies of left main bronchus and those with surgery in their left main bronchus) were excluded from the study. After premedication with 0.03 mg/kg midazolam, 1-2 µg/kg fentanyl and induction of general anesthesia using propofol 1.5 mg/kg and cisatracurium 0.2 mg/kg, a left Robertshaw DLT was inserted. The size of double-lumen tube was selected 0 cm 35F, women > 160 cm 37F; men ≤ 170 cm 39F, > 170 cm 41F. The amount of insertion of double-lumen tube was based on this formula: DLT depth [cm] = 12.5+0.1 × height [cm]). All cases were intubated by three experienced anesthesiologists in the field of thoracic anesthesia using conventional maneuvers and the patients’ height was used to guide the depth of DLT insertion and thereafter the depth of DLT insertion and the position of DLT were checked by auscultation (9). Then flexible bronchoscopy was performed using a 2.8 mm Diameter Olympus video bronchoscope to recheck and confirm the position of DLT or reposition the tube according to the following criteria proposed (10). After positioning the patient to the lateral decubitus, the DLT position was again checked primarily by auscultation and then by Fiber-Optic Bronchoscopy (FOB). Bronchoscopic criteria for evaluating DLT position are as follows: 1- Optimal: The bronchial cuff of DLT is immediately below the tracheal carina, and there is a clear view of the left subcarinal with unobstructed left upper and lower bronchi. 2- Misplacement: Partial dislocation of DLT either proximally or distally. DLT has to be moved (in or out) for more than 0.5 cm to be in the correct position. 3- Critical malposition: DLT dislocated in the trachea or in the right bronchi. Clinical evaluation of the patients and auscultation assessment of DLT position was conducted by three expert anesthesiologists and the endoscopist was not informed of the previous result. Before and after changing the position to the lateral decubitus, auscultation and fiberoptic evaluation were performed by a different anesthesiologist in every case, thus we kept the evaluators blinded. After the bronchoscopic assessment, the DLT was either secured in its position or repositioned according to the bronchoscopic findings. Following a pilot study with 10 cases, we applied nQuery advisor software to calculate the sample size in this agreement study with cross-sectional method. The result showed that 66 cases should be enrolled considering α = 0.05 and β = 0.2. Patients were enrolled in the study with simple sampling method. Statistical analysis using the statistical software package SPSS 16.0 for windows, was performed by application of chi-square test, student’s T-test as well as kappa agreement test. A probability value of less than 0.05 was considered significant. Results Seventy-six adult patients scheduled for elective thoracic surgery, included 33 (43.4%) females, requiring the placement of a left DLT and lateral positioning were enrolled in the study. The demographic data including age (44.8 ± 15.6 years), height (169.5 ± 9.5 cm), weight (72.3 ± 11.9 kg), and body mass index (25.3 ± 3.8 kg/m2) were measured. The DLT sizes used and distribution of right and left thoracotomies were depicted (Table 1). Auscultation and FOB for the first time (supine position) took 34 ± 17 and 86 ± 47 seconds, respectively. Average time for auscultation and FOB at the second time was 30 ± 15 and 74 ± 34 seconds, respectively. After blind intubation, 21.1% of DLTs were considered not to be in the correct position using clinical evaluation. However, bronchoscopy showed that 46.1% of all DLTs were placed incorrectly, 35.5% of them were misplaced, and critical malposition occurred in 10.5% of the patients (Table 2).
Table 1. The size of DLT, lateral decubitus distributions, and Cormack-Lehane grade.
Percent |
Frequency |
Variables Values |
|
22.4 |
17 |
35 |
Size of DLT |
26.3 |
20 |
37 |
|
51.3 |
39 |
39 |
|
53.9 |
41 |
Left |
Lateral decubitus |
46.1 |
35 |
Right |
|
67.1 |
51 |
1 |
Cormack-lehane |
31.6 |
24 |
2 |
|
1.3 |
1 |
3 |
Table 2. Comparison between auscultation and fiberoptic bronchoscopy (FOB) for correct placement of DLT.
Percent |
Frequency |
Position |
Variables |
78.9 |
60 |
Correct |
Auscultation (1st time) |
21.1 |
16 |
Incorrect |
|
53.9 |
41 |
Correct |
FOB (1st time) |
46.1 |
35 |
Incorrect |
|
53.9 |
41 |
Optimal |
FOB (1st time) |
35.5 |
27 |
Misplacement |
|
10.5 |
8 |
Malposition |
|
94.7 |
72 |
Correct |
Auscultation (2nd time) |
5.3 |
4 |
Incorrect |
|
89.5 |
68 |
Correct |
FOB (2nd time) |
10.5 |
8 |
Incorrect |
|
89.5 |
68 |
Optimal |
FOB (2nd time) |
10.5 |
8 |
Misplacement |
Table 3. Correct and incorrect placement of DLT for the 1st & 2nd time
p-value |
Kappa |
FOB 2 |
FOB 1 |
|
Tube size |
||
|
|
Incorrect - |
Correct - |
Incorrect 20(33.3%) |
Correct 40(66.7%) |
|
Auscultation 1 |
0.001 |
|
Correct |
|||||
|
0.421 |
-
4(5.6%) |
-
68(94.4%) |
15(93.3%)
- |
1(6.7%)
- |
Incorrect |
Auscultation 2 |
0.001 |
|
Correct |
|||||
|
0.642 |
4(100%) |
0 |
- |
- |
Incorrect |
|
Table 4. Correct and incorrect placement of DLT for the 1st & 2nd time considering gender.
p-value |
Female n = 33 |
Male n = 43 |
Position of DLT |
|
0.9 |
18(54.5%) |
23(53.5%) |
Correct |
FOB (1st time) |
15(46.5%) |
20(46.5%) |
Incorrect |
||
0.7 |
30(90.9%) |
38(88.4%) |
Correct |
FOB (2nd time) |
3(9.1%) |
5(11.6%) |
Incorrect |
Frequency (percent) is reported.
Table5. Correct and incorrect placement of DLT for the 1st & 2nd time considering the size of DLT.
p-value |
Size of DLT |
Position of DLT |
|||
39Fr n = 39 |
37Fr n = 20 |
35 Fr n = 17 |
|||
p = 0.8 |
20(51.3%) |
11(55%) |
10(58.8%) |
Correct |
FOB (1st time) |
19(48.7%) |
9(45%) |
7(41.2%) |
Incorrect |
||
p= 0.7 |
34(87.2%) |
18(90%) |
16(94.1%) |
Correct |
FOB (2nd time) |
5(12.8%) |
2(10%) |
1(5.9%) |
Incorrect |
Frequency (percent) is reported.
Table 6. The effect of age, height, weight, and BMI on FOB result for the1st time and 2nd time.
FOB (2nd time) Position of DLT
|
FOB (1st time) Position of DLT
|
|||||
p-value |
Incorrect |
Correct |
p-value |
Incorrect |
Correct |
Variables |
0.9 |
49.9±16.4 |
43.8±15.4 |
0.8 |
43.8±15.4 |
45.6±15.9 |
Age (year) |
0.6 |
171.9±10.7 |
170.7±9.9 |
0.9 |
170.7±9.9 |
167.9±9.1 |
Height (cm) |
0.6 |
70.0±11.9 |
71.9±12.9 |
0.6 |
71.9±12.9 |
72.5±11.1 |
Weight (kg) |
0.3 |
23.9±4.8 |
24.7±3.6 |
0.6 |
24.7±3.6 |
25.8±3.9 |
BMI (kg/m2) |
Mean ± SD is reported.
Table 7. Agreement between Auscultation and FOB after blind DLT insertion considering the size of the tube
Kappa
|
Incorrect |
Correct |
Auscultation |
Tube size |
0.161 |
6 (37.5%) |
10 (62.5%) |
Correct |
35 Fr |
|
1 (100%) |
- |
Incorrect |
|
0.239 |
7 (38.9%) |
11 (61.1%) |
Correct |
37 Fr |
|
2 (100%) |
- |
Incorrect |
|
0.586 |
7 (26.9%) |
19 (73.1%) |
Correct |
39 Fr |
|
12 (92.3%) |
1 (7.7%) |
Incorrect |
|
Frequency (percent) is reported.
All the misplaced and malpositioned DLTs were then relocated into the correct position. Following the change in the patients’ position to lateral decubitus, auscultation showed new misplacement in 5.3% of all patients; nevertheless, the fiberoptic assessment revealed DLT misplacement in 10.5% of the patients (Table 2). According to the results of second time FOB, correct positioning of DLT was performed in 68 cases; all had already been diagnosed correctly with auscultation. Despite findings from FOB which showed malposition in 8 patients, auscultation could recognize only 4 cases (50%). Agreement between auscultation and FOB results for the first time was significant (kappa = 0.42) which increased considerably in the second time (kappa = 0.64) (Table 3). Evaluating different variables such as age, sex, height, weight, BMI, and size of DLT, none of them affected incorrect positioning of DLT for the first time. The variables mentioned above and the direction of left or right lateral rotation had no effect on the malposition of DLT (Tables 4-6). In the current study, the kappa coefficient increases with increasing DLT size that means in patients who needed larger DLT size, the greater agreement between auscultation and FOB was recorded after blind intubation (Table 7). Discussion In our study, FOB identified DLT malpositioning in 46.1% of the patients; however, auscultation revealed DLT malpositioning in only 21.1%. Our results are in agreement with previous studies which show that 15.5-44% of left-sided DLTs could not be positioned correctly using clinical signs alone, necessitating fibreoptic bronchoscopy (10,11). Likewise, satisfactory DLT positioning was shown in all cases using clinical criteria; however, subsequent fiberoptic bronchoscopy revealed malposition in 48 % (12). Even some other authors reported a higher misplacement rate of DLT inserted blindly detected by fiberoptic bronchoscopy (13,14). Several other studies showed similar results as far as it was concluded that continuous and routine use of FOB is a must in modern practice (4). A limited number of studies claimed another result. For instance, it was concluded that although 12% of tubes were not in an optimal position, all were in a satisfactory position and non-required repositioning (15). There can be no doubt that the FOB is a useful tool, however, for an experienced anesthetist, this audit has not shown it to be essential. It reported that although bronchoscopy is useful, no double-lumen tube positioning method is fail-safe (6). Operator experience with any method increases the likelihood of success, and a FOB is not always needed for left DLT placement (9). In this study, after turning the patients to lateral decubitus, 10.5% of them required tube repositioning using FOB, but auscultation diagnosed only 5.3% of the malpositions. Possible reasons for this inaccuracy could be that the audibility of breath sounds is often affected by tidal volume, consistency of underlying lung tissue, thickness of the chest wall, noise intensity in the OT, sensitivity of the stethoscope, and hearing acuity of the individual (16). Therefore, according to our findings, final position of DLT should always be verified in the lateral positioning by FOB. Using FOB, tube dislocation secondary to lateral positioning was demonstrated in 27% of their patients (17). Tube misplacement was found in 32% of the patients after lateral positioning (18). Moreover, the study conducted by Desidero et al, showed that there was a significant tracheal movement in 40 of 50 the patients and bronchial movement in 37 of 50 of them (19). The mentioned studies clearly represented that in a significant number of cases, DLT dislocation can occur following the change in the position of patients from supine to lateral decubitus. In our study, evaluating the agreement between FOB and auscultation, we reached kappa = 0.64 for the second time and kappa = 0.42 for the first time. The higher value of kappa in second time may be justified by this fact that DLT placement was corrected prior to lateral positioning. Considering the kappa value, we concluded that in case FOB is not available, auscultation can be helpful as a simple and easy method. In a study carried out by de Bellis et al, there was a complete agreement between the anesthesiologist (auscultation) and the endoscopist in 63% of the patients whose DLT had been correctly placed, as confirmed by flexible bronchoscopy (20). Another conclusion from our study is that the DLT size can influence the agreement between auscultation and FOB; in patients who required larger DLT, the greater agreement between auscultation and FOB was detected. Similar to what was achieved in our study, the blind approach (auscultation) was twice as fast as the directed approach (FOB) (21). According to the study by Boucek et al, although the blind and directed approaches resulted in the successful placement of DLT, more time was required using the directed approach (21). In other studies, such as Cheong et al, the bronchoscope allowed more rapid intubation compared to conventional technique (105 vs. 347 sec) (22). The strong point of the present study was an evaluation of the patients before and after the positioning. None of the patients was deprived of FOB for DLT placement. Conclusion Although we found a high agreement - but not 100% - between fiberoptic bronchoscopy and auscultation for determining the proper placement of DLT, the poorly poisoned DLT can interfere with ideal intraoperative surgical and ventilatory conditions. Thus, DLT position should be checked by fiberoptic bronchoscope after turning the patients to the lateral position ensure that one-lung ventilation and lung isolation would be possible for all the patients. Acknowledgements This study was approval by Tehran University of Medical Sciences- School of Medicine ethics committee with reference number IR.TUMS.REC.1394.903.
5.Cheong KF, Koh KF. Placement of left-sided double-lumen endobronchial tubes: comparison of clinical and fibreoptic-guided placement. Br J Anaesth1999;82(6):920-1.doi:10.1093/bja/ 82.6.920.