Journal of Iranian Medical Council

Journal of Iranian Medical Council

An Analogy between Gray Values of Cone Beam Computed Tomography and Hounsfield Units of CT: A Systematic Review and Meta-Analysis

Document Type : Review article

Authors
Department of Oral Medicine and Radiology, Faculty of Dentistry, Meenakshi Ammal Dental College and Hospital, MAHER University, Chennai-95, India
Abstract
Abstract
Background: This study was conducted to determine and compare the efficacy of Hounsfield Units (HU) of Computed Tomography (CT) and Gray Values (GV) of Cone Beam Computed Tomography (CBCT) in assessing bone Mineral density.
Methods: Literature search was carried out using electronic databases including PubMed, Google scholar, Scopus. In vivo, in vitro and animal studies that analyzed the comparison between the GV of CBCT and HU of CT were included. This review adheres to the Prisma guidelines, and QUADAS-2 tool for risk of bias assessment was performed.
Results: A Total of 4760 studies were roped in for this systematic review, of which 22 articles were included and 8 articles were selected for the meta-analysis. The odds ratio of 8 included articles showed a strong positive correlation between CT and CBCT and the overall classification of 89.86% was obtained. 19 studies had low risk of bias and 4 studies had high risk. Some of the included studies indicated quite low and limited reliability, advocating the need for clinical studies with diagnostic capacity to support the use.
Conclusion: The existing evidence suggests that GVs of CBCT and HUs of CT had a strong positive correlation and the standard formula for the conversion between the two parameters (Gray values and Hounsfield units) need to be derived in future studies with clinical correlation.

Keywords

Subjects


Abstract
Background: This study was conducted to determine and compare the efficacy of Hounsfield Units (HU) of Computed Tomography (CT) and Gray Values (GV) of Cone Beam Computed Tomography (CBCT) in assessing bone Mineral density.
Methods: Literature search was carried out using electronic databases including PubMed, Google scholar, Scopus. In vivo, in vitro and animal studies that analyzed the comparison between the GV of CBCT and HU of CT were included. This review adheres to the Prisma guidelines, and QUADAS-2 tool for risk of bias assessment was performed.
Results: A Total of 4760 studies were roped in for this systematic review, of which 22 articles were included and 8 articles were selected for the meta-analysis. The odds ratio of 8 included articles showed a strong positive correlation between CT and CBCT and the overall classification of 89.86% was obtained. 19 studies had low risk of bias and 4 studies had high risk. Some of the included studies indicated quite low and limited reliability, advocating the need for clinical studies with diagnostic capacity to support the use.
Conclusion: The existing evidence suggests that GVs of CBCT and HUs of CT had a strong positive correlation and the standard formula for the conversion between the two parameters (Gray values and Hounsfield units) need to be derived in future studies with clinical correlation.
Keywords: Animals, Bone density, Cone-beam computed tomography, Laboratory, Reproducibility of results, Search engine, Spiral cone-beam computed tomography, Tomography, X-Ray computed

Introduction
Bone is a connective tissue which alters constantly in living beings (1). Bone constitutes 40% inorganic components, 35% organic components and 25% water (2). Bone remodeling occurs through continuous process of bone resorption and formation, where the net quantity of bone is retained (1). Bone quantity is associated with its mechanical strength, since patients cannot be put through mechanical testing, many non-invasive methodologies have been introduced. Computed Tomography (CT) has been widely used to assess bone density and provides measurement in the form of Hounsfield Units (HU) (3). With the advent of Cone Beam Computed Tomography (CBCT), which provides lesser radiation dose and exposure time along with higher image resolution than conventional CT for evaluating morphologic information. In addition, CBCT has been widely used to determine the Bone density estimation (1). The standard unit for measuring Bone Mineral Density (BMD) in CT is HU. The CT image is displayed as matrix of individual blocks called voxels and each square is called pixel. Each pixel is assigned a CT number which determines the x-ray beam attenuation representing tissue density in the form of arbitrary scale called HU, whereas the ability of CBCT imaging to display differences in photon attenuation is related to the ability of the detector to reveal subtle contrast differences. This parameter is called the bit depth of the system and determines the number of shades of gray available to display the attenuation in the form of Grayscale Values (GV) (4). The main objective of this systematic review is to convert HU to GV and vice versa in measuring bone mineral density and assessing its accuracy in doing so.

Materials and Methods
This systematic review is focused on analyzing the relationship between gray values of CBCT and HU of CT. It follows the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) statement. The PICO guidelines were formatted with:
Population such as animal studies, in vivo and in vitro studies. Intervention used are gray values of CBCT, comparison was done with HU of CT, and the expected outcomes were reliability and analogy between the two parameters.
Inclusion criteria: 
- Animal, in vitro, in vivo studies were taken for the review.
- Studies involving both CT and CBCT imaging modalities were included.
- Only full-text articles in English were included for the study.
- Studies in which correlation coefficient and regression analysis performed were taken for the systematic review.
Exclusion criteria:
- Reviews, personal opinions, studies without reference standards, letters, and conference papers were excluded.
- Articles including other imaging modalities like Dual energy x-ray absorptiometry, Ultrasonography (USG), Magnetic Resonance Imaging (MRI), and Micro CT were excluded.
- Only abstracts, articles with either CT OR CBCT, and other language articles were excluded.
- Studies which lack correlation coefficient, regression analysis were excluded.
The review process involves Study selection, Data extraction, Qualitative assessment, (QUADAS 2), Meta analysis.

Literature search
A literature search was performed using specific strategies in manual and electronic database search using PubMed, ScienceDirect, Google Scholar, and Scopus to identify studies. The Mesh terms used were Multislice Computed Tomography (MSCT) OR Multidetector Computed Tomography (MDCT) AND Cone Beam Computed Tomography OR CBCT AND Correlation AND in vitro studies AND in vivo studies AND animal studies.

Study selection
All the articles were individually reviewed for title and abstract to remove the irrelevant and statistically insignificant ones. Later, full-text articles were retrieved based on the inclusion criteria.

Data extraction 
All the retrieved articles were reviewed individually and data were extracted from each article such as first author name, year, study design, samples used, imaging modality, conversion equations based on the inclusion and exclusion criteria.

Quality assessment
The Quality of studies included in the review were subjected to a risk of bias assessment with the Quality Assessment of Diagnostic Accuracy Studies (QUADAS-2) checklist. The QUADAS-2 tool comprises 4 domains: patient selection, index test, reference standard, and flow and timing, under which signaling questions were included to determine the risk of bias. The results for each item were categorized as yes (Y), unclear (U), or no (N). The summary risk of bias for each study was categorized as low (A), unclear (B), or high (C).

Results
Selection of literature
4760 articles were identified through database and manual search. After removing statistically irrelevant, non- correlated studies and articles involving other imaging modalities 51 articles were screened by reading the abstracts. Ultimately 22 full-text articles were included based on the inclusion and exclusion criteria among which 8 were included for meta-analysis as mentioned in PRISMA Flowchart figure 1.

Study characteristics 
Out of 22 studies included, only 4 were in vivo studies, in which 3 studies were carried out in humans and 1 in rabbits. The remaining 18 were in vitro studies. All 22 studies were done in CT and CBCT. 7 Studies included conversion equation. All the studies evaluated the correlation between CT and CBCT as shown in table 1.

Table 1.  Studies included & its characteristics

Author & Year

Study design

Sample type

Imaging modality

Conversion equation

Magil et al 2017

In-vitro study

Phantom

CT, CBCT

HVL = In (2)
Aluminium
HU material = (material − water)
water x 1000

Nomura et al 2010

In-vitro study

Phantom

MSCT, CBCT

y=0.03669x2+3.602x-350.3 (x: voxel value, y: BMD mg/cm3 HA)

Bastami et al 2017

In-vivo study

5 rabbits-calvaria

CT, CBCT

100 unit increase in GV=112.2

unit increase on HU

Varshowsaz et al 2016

In-vitro study

Phantom

CT, CBCT

NA

Parsa et al 2013

In-vitro study

Human jaws-20

MSCT, CBCT

NA

Chindasombatjaroen et al 2011

In-vitro study

Phantom

MDCT, CBCT

y = 2.0175x + 584.62;

x = pixel value in CBCT; y = CT value in MDCT

Naitoh et al 2009

In-vivo study

16 patients

CT, CBCT

NA

cassetta et al 2013

In-vitro study

20 dry mandibles

CT, CBCT

0.7 X Values of CBCT =
Values of CT

Patrick et al 2017

In-vitro study

20 dry mandibles

MSCT, CBCT

NA

bujtar et al 2014

In-vitro study

human cadaver

CBCT, MSCT

CBCT GV = A X MSCT HU +B

A,B= correlation coefficients

Shokri et al 2018

In-vitro study

Phantom

MDCT, CBCT

NA

Razi et al 2014

In-vitro study

Sheep Head

CT, CBCT

HU=14.621+1.088×gray scale

HU= -24.052+1.146×gray scale

HU= -61.098+1.178×gray scale

Silva et al 2012

In-vitro study

20 mandibles

CBCT, MSCT

NA

sedeek et al 2019

In-vitro study

Phantom

CBCT, MSCT

y=0.682(x)-161 
y=BMD,x=CBCT GV

Khavidet al 2021

In-vitro study

Phantom 52 specimens,
cow rib bone

MDCT, CBCT

NA

Razi et al 2019

In-vivo study

21 patients

CT, CBCT

CBCT=126.92+0.93*CT

Arisan et al 2012

In-vivo study

18 patients

CBCT, MSCT

NA

Gaur et al 2022

In-vitro study

20 goat heads

CT, CBCT

CT mean= 82.3+0.4927CBCT mean

Azeredo et al 2013

In-vitro study

Phantom

CT, CBCT

NA

Dings et al 2019

In-vitro study

5 human dry skulls

MDCT, CBCT

NA

Lee et al 2021

In-vitro study

A dry mandible

CBCT, MSCT

NA

sedeek et al 2019

In-vitro study

Phantom

CBCT, MSCT

y=0.682(x)-161 
y=BMD,x=CBCT GV

Qualitative assessment
The studies were subjected for risk of bias assessment with the help of QUADAS-2 tool. It comprises 4 domains such as patient selection, index test, reference standard, and flow and timing, with 9 signaling questions which helps to judge the study in terms of high, low and unclear risk of bias. For risk of bias assessment, 15 studies (Bujtar et al (5), Bastami et al (6), Cassetta et al (7), Varshowsaz et al (8), Parsa et al (9), Chindasombatjaroen et al (10), Mah et al (11), Razi et al (12), Patrick et al (13), Shokri et al (14), Nomura et al (15), Sedeek et al (16), Gaur et al (17), Dings et al (18), Razi et al (19) had low risk of bias, 4 studies [Silva et al (20), Arisan et al (21), Khavid et al (22), and Lee et al (23)] had high risk of bias, 3 studies [Naitoh et al (24), Azeredo et al (250, Magill et al (26)] had unclear risk of bias.

Meta-analysis
Among 22 studies, 8 studies were included for meta-analysis due to heterogeneity of data in the remaining studies using RevMann software version 5.3. The forest plot analysis was produced between gray value of cone beam computed tomography and HU of computed tomograms (Table 2). Based on the analysis performed using random effects model with inverse variance method, summarized odds ratio was found out. The summarized odds ratio (OR) was 1.08 with a 95% confidence interval of 1.02-1.14 and this was found to be statistically significant as the test for overall effect shows a significance at p<0.05. Heterogeneity of the included eight studies was 0 (I2=0) which confirms the absence of notable variability between the studies. The effect sizes determined across cohorts were uniform in both size and direction as shown in figure 2.
The funnel plot indicates no potential publication bias. The Eggers’ test does not support the presence of funnel plot asymmetry (intercept: -0.39, 95%CI:
-1.22-0.43, t:-0.938, p-value: 0.384) as shown in figure 3.
The odds ratio indicated a strong positive correlation between CT and CBCT with an overall classification of 89.87%.

Table 2. Forest plot analysis showing Odds ratio

 

OR

UPPER 95%CI

LOWER 95%CI

Chindasombatjaroen et al, 2011

1.19

1.32

0.94

Cassetta et al, 2013

1.08

1.23

0.96

Parsa et al, 2013

1.04

1.25

0.83

Razi et al, 2014

1.16

1.31

0.85

Bujtar et al, 2014

1.02

1.19

0.96

Bastami et al, 2017

1.16

1.34

0.8

Razi et al, 2019

1.08

1.28

0.91

Gaur et al, 2022

1.12

1.3

0.94

Discussion
Assessment of bone density serve multiple purposes in dental procedures, including the placement of dental implants and orthodontic micro implants. Additionally, this information proves beneficial in diagnosing conditions like tooth ankylosis, periodontal and endodontic lesions, and in predicting growth patterns and potential. While MSCT is indeed the appropriate instrument for gauging bone density, its widespread adoption in dentistry is hindered by its elevated cost and the substantial radiation dosage it entails (23).
In contrast, CBCT is favored in the dental field due to its lower radiation exposure, brief measurement time, cost-effectiveness, and the relatively high resolution of the images it produces (23). Making a direct comparison of gray density values derived from different CBCTs poses challenges. Unlike MSCT, the attenuation coefficient in CBCT lacks standardization. The gray density values among CBCT scanners are affected by technical variables, including the X-ray beam’s hardening effect, radiation scatter, and the impacts associated with discontinuity in projection data (27).
According to Naitoh et al (24) in 2009 who has done a study with 16 patients, found a high level of correlation between CBCT & MSCT(r=0.965). Voxel values from Mandible cancellous bone was used for estimating BMD. Razi et al (12), performed a study with 21 patients (16 Males & 5 Females) totally in 25 soft and hard tissues, found strong correlation between CT & CBCT (R2 =0.85,0.74), respectively. According to Shokri et al (14), the size of FOV significantly changed mean gray values of bone substitutes, whereas CBCT with small FOV had significant correlation with MDCT results. Razi et al (19) performed a study comparing the GV of CBCT with HU of CT with 3 different types of CBCT scanners and 1 CT scanner and derived 3 linear regression equations which revealed a strong correlation between GV and HU.
A systematic review conducted by Eugren et al (28) in 2022 which included only 3 articles to conclude that GV of CBCT cannot be correlated to the HU of CT which was attributed to lack of clinical studies with diagnostic capacity. On the contrary, the current study, with 8 articles provided a strong positive correlation, in which Bujtar et al (5) and Parsa et al (9) showed the highest level of linear correlation.

Limitations and future perspectives
Lack of homogeneity of data in the included studies is one of the major drawbacks of this systematic review. More clinical studies need to be incorporated to derive a standard conversion formula between HU and GV irrespective of the type of scanner used.

Conclusion
This systematic review illustrated that converting GV to HU involves both qualitative and quantitative assessments. Thus, Gray values obtained can be used for estimation of bone mineral density with the proper conversion formula. However, standardization of equipment calibration, correlation methods, and conversion equations is necessary, regardless of the software utilized.

Acknowledgement
The study has been approved by the Ethical committee of Meenakshi Ammal Dental College & Hospital, MAHER University, Chennai.

Conflict of Interest 
Authors declare no conflict of interest.

  1. References

    1. Kim DG. Can dental cone beam computed tomography assess bone mineral density? J Bone Metab 2014 May 31;21(2):117-26. https://pubmed.ncbi.nlm.nih.gov/25006568/
    2. Morgan EF, Barnes GL, Einhorn TA. Osteoporosis. San Diego: Academic Press; 2008. pp. 3-26.
    3. Genant HK, Engelke K, Fuerst T, Glüer CC, Grampp S, Harris ST, et al. Noninvasive assessment of bone mineral and structure: state of the art. J Bone Miner Res 1996;11:707-30. https://pubmed.ncbi.nlm.nih.gov/8725168/
    4. Mallya S, Lam E. White and Pharoah’s oral radiology, principles and interpretation. 8th ed. Mosby; 2018. 672 p.
    5. Bujtár P, Simonovics J, Zombori G, Fejer Z, Szucs A, Bojtos A, et al. Internal or in-scan validation: a method to assess CBCT and MSCT gray scales using a human cadaver. Oral Surg Oral Med Oral Pathol Oral Radiol 2014;117:768-79. https://pubmed.ncbi.nlm.nih.gov/24842449/
    6. Bastami F, Shahab S, Parsa A, Abbas FM, Noori Kooshki MH, Namdari M, et al. Can gray values derived from CT and cone beam CT estimate new bone formation? An in vivo study. Oral Maxillofac Surg 2018 Mar;22:13-20. https://pubmed.ncbi.nlm.nih.gov/29086089/
    7. Cassetta M, Stefanelli LV, Pacifici A, Pacifici L, Barbato E. How accurate is CBCT in measuring bone density? A comparative CBCT-CT in vitro study. Clin Implant Dent Relat Res 2014;16(4):471-8. https://pubmed.ncbi.nlm.nih.gov/23294461/
    8. Varshowsaz M, Goorang S, Ehsani S, Azizi Z, Rahimian S. Comparison of tissue density in Hounsfield units in computed tomography and cone beam computed tomography. J Dent (Tehran) 2016;13:108-15. https://pubmed.ncbi.nlm.nih.gov/27928239/
    9. Parsa A, Ibrahim N, Hassan B, van der Stelt P, Wismeijer D. Bone quality evaluation at dental implant site using multislice CT, micro-CT and cone beam CT. Clin Oral Implants Res 2015;26(1):e1-7. https://pubmed.ncbi.nlm.nih.gov/24325572/
    10. Chindasombatjaroen J, Kakimoto N, Shimamoto H, Murakami S, Furukawa S. Correlation between pixel values in a conebeam computed tomographic scanner and the computed tomographic values in a multidetector row computed tomographic scanner. J Comput Assist Tomogr 2011;35:662-5. https://pubmed.ncbi.nlm.nih.gov/21926866/
    11. Mah P, Reeves TE, McDavid WD. Deriving Hounsfield units using grey levels in cone beam computed tomography. Dentomaxillofac Radiol 2010;39:323-3. https://pubmed.ncbi.nlm.nih.gov/20729181/
    12. Razi T, Emamverdizadeh P, Nilavar N, Razi S. Comparison of the Hounsfield unit in CT scan with the gray level in cone-beam CT. J Dent Res Dent Clin Dent Prospects 2019;13:177-82. https://pubmed.ncbi.nlm.nih.gov/31857863/
    13. Patrick S, Birur NP, Gurushanth K, Raghavan AS, Gurudath S. Comparison of gray values of cone-beam computed tomography with Hounsfield units of multislice computed tomography: an in vitro study. Indian J Dent Res 2017;28:66-70. https://pubmed.ncbi.nlm.nih.gov/28393820/
    14. Shokri A, Ramezani L, Bidgoli M, Akbarzadeh M, Ghazikhanlu-Sani K, Fallahi-Sichani H. Effect of field-of-view size on gray values derived from cone-beam computed tomography compared with the Hounsfield unit values from multidetector computed tomography scans. Imaging Sci Dent 2018 Mar 1;48(1):31-9. https://pubmed.ncbi.nlm.nih.gov/29581947/
    15. Nomura Y, Watanabe H, Honda E, Kurabayashi T. Reliability of voxel values from cone-beam computed tomography for dental use in evaluating bone mineral density. Clin Oral Implants Res 2010;21:558-62. https://pubmed.ncbi.nlm.nih.gov/20443807/
    16. A Sedeek H, A El-Awady A, K Mohamed M. Bone density assessments in multislice and cone-beam computed tomography using water, plaster of paris and motor oil phantom. Al-Azhar J Dent Sci 2019 Jan 1;22(1):95-101. https://ajdsm.journals.ekb.eg/article_71728.html
    17. Gaur A, Dhillon M, Puri N, Sethi Ahuja U, Rathore A. Questionable accuracy of CBCT in determining bone density: A comparative CBCT–CT in vitro study. Dent Med Probl 2022;59(3):413-19. https://pubmed.ncbi.nlm.nih.gov/36196514/
    18. Dings J, Paulus J, Verhamme L, Merkx MA, Xi T, Meijer GJ, et al. Reliability and accuracy of cone beam computed tomography versus conventional multidetector computed tomography for image-guided craniofacial implant planning: An in vitro study. Int J Oral Maxillofac Implants 2019 May 1;34(3). https://pubmed.ncbi.nlm.nih.gov/30934042/
    19. Razi T, Niknami M, Alavi Ghazani F. Relationship between Hounsfield unit in CT scan and gray scale in CBCT. J Dent Res Dent Clin Dent Prospects 2014;8:107-10. https://pubmed.ncbi.nlm.nih.gov/25093055/
    20. Silva IM, Freitas DQ, Ambrosano GM, Bóscolo FN, Almeida SM. Bone density: comparative evaluation of Hounsfield units in multislice and cone-beam computed tomography. Braz Oral Res 2012 Nov-Dec;26(6):550-6. https://pubmed.ncbi.nlm.nih.gov/23184166/
    21. Arisan V, Karabuda ZC, Avsever H, Özdemir T. Conventional multi-slice computed tomography (CT) and cone-beam CT (CBCT) for computer-assisted implant placement. Part I: relationship of radiographic gray density and implant stability. Clin Implant Dent Relat Res 2013 Dec;15(6):893-906. https://pubmed.ncbi.nlm.nih.gov/22251553/
    22. Khavid A, Sametzadeh M, Godiny M, Moarrefpour MM. Comparison of the Hounsfield unit values obtained from cone-beam computed tomography (CBCT) and multidetector computed tomography (MDCT) images for different bone densities. J Contemp Med Sci 2021 Mar;7(2):92-5. https://www.jocms.org/index.php/jcms/article/view/943
    23. Lee MY, Park JH, Chang NY, Seo HY, Chae JM. Bone density measurements: multi-slice computed tomography versus cone-beam computed tomography. Clin J Korean Assoc Orthodon 2021 Mar;11(1):1-0. DOI: 10.33777/cjkao.2021.11.1.1
    24. Naitoh M, Hirukawa A, Katsumata A, Ariji E. Evaluation of voxel values in mandibular cancellous bone: relationship between cone-beam computed tomography and multislice helical computed tomography. Clin Oral Implants Res 2009 May;20(5):503-6. https://pubmed.ncbi.nlm.nih.gov/19250241/

    25.Azeredo F, de Menezes LM, Enciso R, Weissheimer A, de Oliveira RB. Computed gray levels in multislice and cone-beam computed tomography. Am J Orthod Dentofacial Orthop 2013;144:147-55. https://pubmed.ncbi.nlm.nih.gov/23810056/

    26.Magill D, Beckmann N, Felice MA, Yoo T, Luo M, Mupparapu M. Investigation of dental cone-beam CT pixel data and a modified method for conversion to Hounsfield unit (HU). Dentomaxillofac Radiol 2017;47(2):20170321. https://pubmed.ncbi.nlm.nih.gov/29076750/

    27.Pauwels R, Jacobs R, Singer SR, Mupparapu M. CBCTbased bone quality assessment: are Hounsfield units applicable? Dentomaxillofac Radiol 2015;44:201402. https://pubmed.ncbi.nlm.nih.gov/25315442/

    28.Eguren M, Holguin A, Diaz K, Vidalon J, Linan C, PachecoPereira C, et al. Can gray values be converted to Hounsfield units? A systematic review. Dentomaxillofac Radiol 2022;51:20210140. https://pubmed.ncbi.nlm.nih.gov/34148350/